Ask An Expert

Mini Centrifuge LMI-A24 Catalog

Mini Centrifuge LMI-A24

Mini Centrifuge LMI-A24

Overview

Mini Centrifuge LMI-A24 is a handheld centrifuge with Centrifugal speed of 4,000 rpm. ± 3 % and maximum RCF of 990 x g it is equipped with brushless DC motor gives noiseless and vibration free operation mode. It has ergonomic compact design and user friendly usage mode. It can fit with 2 ml, 1.5 ml, 0.5 ml and 0.2 ml tubes. Ideal for macromolecules separation and clinical, PCR and HPLC analysis protocols. The handheld centrifuge is very suitable for microtubule filtration and fast centrifugation, the micro blood cell separation, microbial sample processing, experimental partition centrifugal of PCR.

Features

  • - Enhanced flash memory microcontroller provides high performance
    Uses less bench area space
    Model can fit in 2 ml, 1.5 ml, 0.5 ml and 0.2 ml tubes
    Light-weight, small size and beautiful design
    Ambient Temperature required is 23 °C
    High speed safety feature protect from unit damage
    Compact design gives it a tool-free rotor exchange feature
    less noise and vibration free operation mode
    Visual alarm and false audio system
    Shock absorbent RSS material represent more stable operations
    Interlock lid prevents tubes exclusion during spinning
    User friendly usage and easy to install
    No time control button, unless you open the lid, then it stop centrifugation
Specifications
Maximum Speed 7000 rpm ± 3 %
Maximum RCF 2910 x g
Allowable ambient temperature 5 to 40 °C
Relative Humidity 80 % RH
Fastest acceleration time ≤ 12 s
Fastest Deceleration time ≤ 16 s
Motor DC 24 V
The noise at highest speed ≤ 43 dB (A)
Rate of work Input 18 W
Power Supply AC85-264V/50/60Hz
Power Fuse PPTC / self-recovery fuse / No replacement needed
Control Method PI / Power direct drive
Protection Level IP 21
Rotor Capacity 2 / 1.5 / 0.5 / 0.2 ml × 80.2 ml × 8 × 4 (PCR)
PackagingDimension 280×250×180 mm
Packaging Weight 2 kg
Standard Accessories :

Standard Acc. No. Rotor Name Measurement
1 Angle Rotor 01 8×1.5/2 ml angle
2 PCR Rotor 01 0.2 ml ×8×4
3 Adatper 01 0.5 ml
4 Adatper 02 0.2 ml

Applications :

Mini Centrifuge is widely used in the extraction of nucleic acids, cell debris separation and microfiltration from the biological samples. It is preferably used for protein concentration, PCR analyses, clinical analysis and HPLC protocols.

Enquiry Now