Ask An Expert
Air Jacketed CO 2 Incubator LAJI-C20 is a shaking CO2 incubator with air jacketed configurations suitable for heat/UV sterilization and temperature range of 4˚ to 65˚C with shaking frequency of 40 to 300 rpm. Designed with CO2 concentration sensor, advanced alarm system, temperature control and monitoring system, it provides uniform and stable conditions. Incorporated 90˚ C high temperature sterilization method, touch screen with PID controller, and enhanced safety features provides the benefits of contamination control and uncompromising temperature uniformity for even the most demanding incubations.
CO2 range | 0-20% |
CO2 recovery time | (Return to 5% after opening the door for 30 seconds) ≤3 mins |
CO2 control accuracy | ±0.1% |
CO2 stability of 5%CO2 | ±0.1% |
CO2 uniformity | ±0.1% |
CO2 sensor | IR CO2 sensor |
Shaking frequency | 40-300 rpm |
Shaking Precision | ±1 rpm |
Amplitude | 20 mm |
Drive way | Three eccentric shaft balance drive type |
Humidity control | 50 to 90% |
37C humidity control accuracy | The chamber can reach 85%-95% when 37˚ C , and the humidity can be controlled in real time |
Humidity resolution | 0.001 |
Humidity deviation | ±2% |
Related humidity | Nature vaporate > 95% |
Allowable shaking load | 10 kg |
Temperature control range | 4-65° C |
Temperature control accuracy | ±0.1° C |
Temperature uniformity | ±0.5° C |
Temperature sensor | PT 100 |
Environment temperature detector | Independent |
Over temperature protection system | Independent |
Temperature recover time | (Return to 37˚ C after opening the door for 30 sec) ≤8 mins |
Heating method | 6 side heating |
Disinfection method | 90˚ C high temp |
Timing range | 0 to 99 hour 59 min |
Platform size (L×B) | 350×350 mm |
Shelf | 1 |
Display | Digital Color display with touch panel |
Sterilization system | 90 degree high temp. high humidity sterilization system (RHP); UV sterilization (option) |
Controller | PID |
Interface | RS485 |
Door Open/close message | Yes |
Alarm system | High and low temperature and over temperature; Chamber sensor failure; Independent temperature limiter; Door open; High/low CO2 condensation; Power off alarm; Door temperature sensor failure; Disinfection and sterilization status reminder; Over temp sensor failure and Abnormal function |
Interior dimension (W×D×H) | 505×600×420 mm |
Exterior dimension (W×D×H) | 693×875×753 mm |
Gross Weight | 91 kg |
Power | 1500 W |
Voltage | 220V 50Hz |
Flask Configuration | 50 ml×36, 100 ml×23, 250 ml×13, 500 ml×8, 1 L×5, 2 L×3, 4 L×1, 5 L×1 |
Air Jacketed CO2 Incubator are used for tissue engineering, cell culture, in vitro fertilization, diagnostics, neuroscience, cancer research and for developing biosensors across medical research and the pharmaceutical industry.