Ask An Expert
Liquid Particle Counter LLPC-A10 comes with high precision syringe sampling and semiconductor laser sensor. Designed using photoresist method principle, it detects size and number of solid particles in a liquid sample. It’s a high quality performance instrument, designed in accordance with international standards such as ISO4406:1999.
Measurement range | 1 to 250 µm (25 sensors) 1 µm to 320 µm (32 sensors) 2 µm to 400 µm (40 sensors) |
Light Source | Semiconductor laser light |
Measurement Channels | 8 channels |
Particle size | 1 µm (2 µm) to arbitrarily set within 100 µm |
Sensitivity | 1 µm (ISO04402) 4 µm (C) (ISO11171) (25, 32 sensor) 2 µm (ISO4402) (40 sensor) |
Sample volume | ≥ 0.5 ml |
Sample accuracy | Better than ± 1 % |
Sampling rate | 5 ml/ min to 60 ml/ min |
Resolution | < 10 % |
Coincidence error limit | 10,000/ ml (5 % Coincidence error) |
Maximum cabin pressure | 0.6 MPa |
Maximum cabin vacuum | 0.08 MPa |
Maximum viscosity detection | 350 cps (40 sensor) 250 cps (32 sensor) 100 cps (25 sensor) |
Ambient temperature | 10 to 40°C |
Interface | RS232 |
Internal Dimension | 430 × 320 × 590 mm |
Internal Weight | 55 kg |
External Packaging Dimension | 450 × 340 × 620 mm |
External Packaging Weight | 75 kg |
Power | 80 W |
Power supply | 220 V, 50 Hz |
Used in petroleum, metallurgy, automobile, aviation, chemical and other fields to detect insoluble particles present in hydraulic oil, petroleum oils, turbine and insulating oils, liquid polymer solution.