Overview

Laser Particle Size Analyzer LLPA-C10 is a fully automatic unit integrated with wet and dry dispersion. With Laser diffraction principle and Mie and Fraunhofer scattering, features wet measuring range of 0.01 to 2000 µm and dry measuring range of 0.1 to 2000 µm. Designed with semiconductor lasers, log-spaced array detector and silicon photodetectors, offers full-automatic and manual operation mode with automatic optical path alignment system. Built-in sample dispersion system and unconstrained free fitting technology comprised software, provides high precision data with <2 min test time for wet dispersion and <1 min test time for dry dispersion. With different analysis mode, statistic method and comparison, it supports intelligent operation, user-defined analysis and test reports in various format.

Features

  • - A fully automatic unit integrated with wet and dry dispersion, convenient operation
    Laser diffraction principle and Mie and Fraunhofer scattering, rapid measurement
    Wet measuring range of 0.01 to 2000 µm and dry measuring range of 0.1 to 2000 µm
    Semiconductor lasers with log-spaced array detector and silicon photodetectors
    Full-automatic and manual operation mode with automatic optical path alignment system
    Built-in sample dispersion system, ensures sufficient dispersion and accurate result
    Unconstrained free fitting technology comprised software, offers high precision data
    <2 min test time for wet dispersion and <1 min test time for dry dispersion
    Different analysis mode, statistic method and comparison, meet different requirements
    Supports intelligent operation, user-defined analysis and test reports in various format
    Designed as per ISO 13320-1:2009 standard configurations
    High-efficient, stable and reliable unit for accurate and repeatable results
Specifications
Dispersion TypeWet and Dry
Measurement RangeWet: 0.01 to 2000 µm Dry: 0.1 to 2000 µm
PhotodetectorsSilicon photodetectors
Channel numberWet: 127 Dry: 100
Accuracy ErrorWet and Dry:
Repeatability ErrorWet and Dry:
Light SourceHigh performance semiconductor laser( λ= 632.8nm P >2.0 mW) Auxiliary blue semiconductor laser (λ= 405 nm, P >2.0 mW)
DetectorLog-spaced array Test angle from 0.015 degree to 145 degree
Wet DispersionUltrasonic Frequency: 40 kHz
Ultrasonic Time: ≥1 S
Ultrasonic Power: 60 W
Stir Revolution speed: 0 to 300 rpm (Adjustable)
Circulation: Rated Flow: 30 L/min Rated Power: 70 W
Sample tank: 1000 ml
Dry DispersionDry-turbulence dispersion Normal shock wave shear technique
Optical Path AlignmentFully automatic
Operation modeAutomatic/ Manual (Switchable)
StandardISO 13320-1:2009
Test SpeedWet:
Software RunningWIN 7/10/XP, 64 bits
Analysis modeFree Distribution, R-R Distribution, Logarithm Normal Distribution, Mesh number classification etc.
Statistic MethodVolume Distribution, Quantity Distribution
Test ReportWord, Excel, Photo( Bmp), Text etc.
Intelligent operationIntelligent SOP operation
Dimension (L×W×H)1040×440×540 mm
Net Weight70 kg
Optional Accessories :

10 ml micro sample pool

Applications :

Laser particle size analyzer is used to measure the sizes of particles in a material across cosmetics, food, coal, paint, paper, petrochemical, dyes, inks, explosives, , kaolin, medicine, metal powder, mica, milling, minerals, oxides industries etc.