Ask An Expert
Disintegration tester LDIT-A11 is adopted with high-precision digital electronic sensor, ensures quality precision and accuracy of the water bath temperature without need for calibration. Equipped with two to three sets of basket, conducts independent disintegration experiments. The default setting for lifting time of the basket is 15 mins, can manually set the other time value as per the experimental need.
Number of basket lifting set | 2 sets |
Lifting frequency of the basket | 30 to 32/ min |
Lifting distance of basket | 55 ± 1 mm |
Minimum spacing from screen to bottom of beaker | 25 ± 2 mm |
Mesh size | 2 mm (as per standard configuration ) |
Automatic time range | 5 to 120 min |
Work noise | < 60 dB |
Temperature prefabrication range | Room Temperature to 50°C, display resolution: 0.1 °C |
Temperature control accuracy | ± 0.3°C |
Time prefabrication range | 1 to 900 min |
Time control accuracy | 0.5 min |
Accuracy of temperature control | 37 ± 0.5°C |
Dimension (L × W × H) | 400 × 320 × 440 mm |
Power supply | 220 V/50 Hz |
Total power | 600 W |
Gross Weight | 20 kg |
Accessories no. | Name | Capacity | Qty. |
1 | Beaker | 1000 ml | 2 |
Accessories no. | Name |
1 | 0.425 mm diameter mesh |
2 | 1 mm diameter mesh |
3 | Beaker |
Used for detection of rate of disintegration of solids such as tablets, capsules, pills in required /set (prescriptive) conditions at pharmaceutical industry.