Double Beam UV-Vis Spectrophotometer LUS-B13 is a compact, tabletop double beam designed UV-Vis spectrophotometer comprising Silicon photodiode detector, Tungsten and Deuterium lamp as light source. Equipped with wide wavelength range of 190 nm to 1100 nm, it offers 1.0 nm of bandwidth with automatic setting of wavelength. Designed with 6 inches LCD display, import source and receiver system, it provides rich measurement methods with high performance and reliability.
Wavelength Range | 190 to 1100 nm |
Spectral Bandwidth | 1.0 nm |
Optical System | Double beam |
Wavelength Accuracy | ±0.3 nm |
Wavelength Repeatability | ≤0.1 nm |
Wavelength Resolution | 0.1 nm |
Photometric Display Range | -4 to 4 A |
Photometric Mode | T, A, C, E |
Photometric Accuracy | ±0.3% τ (0 to 100%τ) ± 0.002 A (0 to 0.5 A) ± 0.003 A (0.5 A to 1 A) |
Photometric Repeatability | 0.15% τ (0 to 100% τ) ± 0.001 A (0 to 0.5 A) ± 0.0015 A (0.5 A to 1 A) |
Light Source | Tungsten lamp, Deuterium lamp |
Stray Light | ≤0.03% τ (220 nm NaI, 340 nm NaNO2) |
Stability | 0.0005 A/h @500 nm |
Noise | ± 0.002 A @500 nm |
Display | 6 inches high brightness Blue LCD display |
Baseline Flatness | ± 0.001 A |
Detector | Silicon photodiode |
Power Supply | AC 220/110 V, 50/60 Hz |
Dimensions (L×W×H) | 560×450×230 mm |
Net Weight | 28 kg |
Double Beam UV-Visible Spectrophotometer are used for analysis of band gap, optical coatings and thin films, quantitative analyses, kinetics, wavelength scanning, and DNA and Protein analysis across biological research, bio-industry, pharmaceutical analysis, pharmaceutical, teaching and research, environmental protection, food hygiene, clinical examination, health and epidemic prevention and other fields.