Salt Spray Test Chamber LSSC-A15

Salt Spray Test Chamber LSSC-A15

Overview

Salt Spray Test Chamber LSSC-A15 is designed as per ASTMB117, ISO9227 configurations, and is comprised with 1600 L of capacity and external 78 L salt solution tank with a brine mixing system. Features continuous/periodic spray with adjustable spray tower and quartz nozzle with high temperature, corrosion and clogging resistance. Incorporated with PID programmable touch screen controller, PT100 Ω / MV A-class sensor, Nichrome heater and Saturated air barrel, this salt spray test chamber has air source diverting device to distribute the compressed air to the opening system. With round bars and V-shaped groove sample holder, it is provided with protection against humidifier dry-combustion, over-temperature, over-current, water shortage and earth leakage. Designed with glass fiber reinforced plastics, transparent window, pneumatic switch lid, pneumatic seal and adjustable castors with brakes, this salt spray test chamber ensures long service life with high temperature and corrosion resistance.

Features

  • - Designed as per ASTMB117, ISO9227 configurations
    1600 L of capacity with external 78 L salt solution tank
    Continuous/periodic spray with adjustable spray tower and Quartz nozzles
    Quartz nozzle with high temperature, corrosion and clogging resistance
    PID programmable touch screen controller with PT100 Ω / MV A-class sensor
    Nichrome heater as heating element with Saturated air barrel
    Air source diverting device to distribute the compressed air to the opening system
    Round bars and V-shaped groove sample holder for different tests
    Protection against humidifier dry-combustion, over-temperature, over-current, water shortage and earth leakage
    Glass fiber reinforced plastics, transparent window, pneumatic switch lid and seal
    Adjustable castors with brakes for easy and convenient movement
    Long service life with high temperature and corrosion resistance
Specifications
Testing room capacity 1600 L
Temperature range Ambient to +60°C
Humidity range 95% to 98% RH
Temperature Fluctuation ± 0.5 ℃
Temperature Deviation ± 2.0 ℃
Spray deposition 1 to 2 ml / 80 cm2 .h
Spray type Continuous /Periodic
Spraying system Atomizer tower and Spray nozzle
Air preheating Saturated air barrel; 31 L
Heating element Nichrome heater
Salt fog collection Fog collector and Fog measure cylinder
Standard sample holder 8 round bars and 7 V-shaped grooves
Controller PID
Sensor PT100 Ω / MV A-class sensor
Housing Glass fiber reinforced plastics
Observation window Glass
Sealing Pneumatic seal
Protection Humidifier Dry-combustion Protection;
Over-temperature Protection;
Over-current Protection;
Water Shortage Protection;
Earth leakage Protection
Maximum noise 65 dBA
Power supply 220 V 50 Hz 1 P
Water solution tank dimension (W×D×H) 600×370×350 mm
Observation window dimension (L×B) 800×300 mm
Internal dimension (W×D×H) 2000×1000×800 mm
External dimension (W×D×H) 3000×1420×1850 mm
Net weight 480 kg
Gross weight 630 kg
Applications :

Salt spray test chamber is used to create and maintain the salt spray (fog) test environment, and test the anti-corrosion quality of all the materials surfaces after the rust-proof of painting, coating, electroplating, anodizing and rust-proof of greasing.

Enquiry Now