Analytical HPLC-A7832 system is equipped with a flow rate range of 0.0001 to 10 ml per minute. Integrated with a single and dual wavelength UV-Vis detector with a range of 190 to 800 nm that analyzes sample’s absorption at different wavelengths and a high pressure pump providing continuous consistent flow. Built in column heater provides reproducible retention time.
Used in pharmaceutical analysis, food nutrient analysis of amino acids and sugars, detection of food contaminants, environmental pollutant and chemical product purity analysis
Detector Specifications | Model No HPD-A20 HPD-A21 Wavelength range Single wavelength ( 190 to 800 nm ) Dual wavelength ( 190 to 800 nm ) Wavelength accuracy ± 0.5 nm ± 2 nm Wavelength repeatability 0.15 nm 0.3 nm Band width 8 nm 8 nm Optical path 10 nm 10 nm Deuterium lamp 190 to 400 nm 190 to 400 nm Tungsten lamp 400 to 800 nm 400 to 800 nm Baseline noise ( Static ) Au ≤ ± 0.35 × 10-5 ≤ ± 0.5 × 10-5 Baseline drift ( Static ) Au / h ≤ 1 × 10-4 ≤ 1.5 × 10-4 Minimum detection limit ≤ 4 × 10-8 ≤ 4 × 10-8 Power 100 W 100 W Power supply 90 V / 50 to 60 Hz 90 V / 50 to 60 Hz Dimension 460 × 315 × 180 mm 460 × 315 × 180 mm Weight 9.5 kg 9.5 kg |
Pump Specifications |
Flow rate range: 0.0001 to 10 ml / min Maximum pressure: 50 MPa , 0 to 2 ml / min Column size: 4.6 × 250 mm Sample size: 5 µm Pressure pulsation: ≤ 1 % at 10 MPa Flow rate accuracy: ± 0.15 % at 1 ml / min Flow rate repeatability %: ≤ 0.1 Gradient accuracy %: ± 1.5 Gradient repeatability %: ≤ 0.2 Dimension: 460 × 310 × 180 mm Power supply: 90 V Weight: 11.6 kg |
Detector Specifications | Model No HPD-A20 HPD-A21 Wavelength range Single wavelength ( 190 to 800 nm ) Dual wavelength ( 190 to 800 nm ) Wavelength accuracy ± 0.5 nm ± 2 nm Wavelength repeatability 0.15 nm 0.3 nm Band width 8 nm 8 nm Optical path 10 nm 10 nm Deuterium lamp 190 to 400 nm 190 to 400 nm Tungsten lamp 400 to 800 nm 400 to 800 nm Baseline noise ( Static ) Au ≤ ± 0.35 × 10-5 ≤ ± 0.5 × 10-5 Baseline drift ( Static ) Au / h ≤ 1 × 10-4 ≤ 1.5 × 10-4 Minimum detection limit ≤ 4 × 10-8 ≤ 4 × 10-8 Power 100 W 100 W Power supply 90 V / 50 to 60 Hz 90 V / 50 to 60 Hz Dimension 460 × 315 × 180 mm 460 × 315 × 180 mm Weight 9.5 kg 9.5 kg |
Pump Specifications |
Flow rate range: 0.0001 to 10 ml / min Maximum pressure: 50 MPa , 0 to 2 ml / min Column size: 4.6 × 250 mm Sample size: 5 µm Pressure pulsation: ≤ 1 % at 10 MPa Flow rate accuracy: ± 0.15 % at 1 ml / min Flow rate repeatability %: ≤ 0.1 Gradient accuracy %: ± 1.5 Gradient repeatability %: ≤ 0.2 Dimension: 460 × 310 × 180 mm Power supply: 90 V Weight: 11.6 kg |
Detector Specifications | Model No HPD-A20 HPD-A21 Wavelength range Single wavelength ( 190 to 800 nm ) Dual wavelength ( 190 to 800 nm ) Wavelength accuracy ± 0.5 nm ± 2 nm Wavelength repeatability 0.15 nm 0.3 nm Band width 8 nm 8 nm Optical path 10 nm 10 nm Deuterium lamp 190 to 400 nm 190 to 400 nm Tungsten lamp 400 to 800 nm 400 to 800 nm Baseline noise ( Static ) Au ≤ ± 0.35 × 10-5 ≤ ± 0.5 × 10-5 Baseline drift ( Static ) Au / h ≤ 1 × 10-4 ≤ 1.5 × 10-4 Minimum detection limit ≤ 4 × 10-8 ≤ 4 × 10-8 Power 100 W 100 W Power supply 90 V / 50 to 60 Hz 90 V / 50 to 60 Hz Dimension 460 × 315 × 180 mm 460 × 315 × 180 mm Weight 9.5 kg 9.5 kg |
Pump Specifications |
Flow rate range: 0.0001 to 10 ml / min Maximum pressure: 50 MPa , 0 to 2 ml / min Column size: 4.6 × 250 mm Sample size: 5 µm Pressure pulsation: ≤ 1 % at 10 MPa Flow rate accuracy: ± 0.15 % at 1 ml / min Flow rate repeatability %: ≤ 0.1 Gradient accuracy %: ± 1.5 Gradient repeatability %: ≤ 0.2 Dimension: 460 × 310 × 180 mm Power supply: 90 V Weight: 11.6 kg |
Detector Specifications | Model No HPD-A20 HPD-A21 Wavelength range Single wavelength ( 190 to 800 nm ) Dual wavelength ( 190 to 800 nm ) Wavelength accuracy ± 0.5 nm ± 2 nm Wavelength repeatability 0.15 nm 0.3 nm Band width 8 nm 8 nm Optical path 10 nm 10 nm Deuterium lamp 190 to 400 nm 190 to 400 nm Tungsten lamp 400 to 800 nm 400 to 800 nm Baseline noise ( Static ) Au ≤ ± 0.35 × 10-5 ≤ ± 0.5 × 10-5 Baseline drift ( Static ) Au / h ≤ 1 × 10-4 ≤ 1.5 × 10-4 Minimum detection limit ≤ 4 × 10-8 ≤ 4 × 10-8 Power 100 W 100 W Power supply 90 V / 50 to 60 Hz 90 V / 50 to 60 Hz Dimension 460 × 315 × 180 mm 460 × 315 × 180 mm Weight 9.5 kg 9.5 kg |
Pump Specifications |
Flow rate range: 0.0001 to 10 ml / min Maximum pressure: 50 MPa , 0 to 2 ml / min Column size: 4.6 × 250 mm Sample size: 5 µm Pressure pulsation: ≤ 1 % at 10 MPa Flow rate accuracy: ± 0.15 % at 1 ml / min Flow rate repeatability %: ≤ 0.1 Gradient accuracy %: ± 1.5 Gradient repeatability %: ≤ 0.2 Dimension: 460 × 310 × 180 mm Power supply: 90 V Weight: 11.6 kg |
Detector Specifications | Model No HPD-A20 HPD-A21 Wavelength range Single wavelength ( 190 to 800 nm ) Dual wavelength ( 190 to 800 nm ) Wavelength accuracy ± 0.5 nm ± 2 nm Wavelength repeatability 0.15 nm 0.3 nm Band width 8 nm 8 nm Optical path 10 nm 10 nm Deuterium lamp 190 to 400 nm 190 to 400 nm Tungsten lamp 400 to 800 nm 400 to 800 nm Baseline noise ( Static ) Au ≤ ± 0.35 × 10-5 ≤ ± 0.5 × 10-5 Baseline drift ( Static ) Au / h ≤ 1 × 10-4 ≤ 1.5 × 10-4 Minimum detection limit ≤ 4 × 10-8 ≤ 4 × 10-8 Power 100 W 100 W Power supply 90 V / 50 to 60 Hz 90 V / 50 to 60 Hz Dimension 460 × 315 × 180 mm 460 × 315 × 180 mm Weight 9.5 kg 9.5 kg |
Pump Specifications |
Flow rate range: 0.0001 to 10 ml / min Maximum pressure: 50 MPa , 0 to 2 ml / min Column size: 4.6 × 250 mm Sample size: 5 µm Pressure pulsation: ≤ 1 % at 10 MPa Flow rate accuracy: ± 0.15 % at 1 ml / min Flow rate repeatability %: ≤ 0.1 Gradient accuracy %: ± 1.5 Gradient repeatability %: ≤ 0.2 Dimension: 460 × 310 × 180 mm Power supply: 90 V Weight: 11.6 kg |
Detector Specifications | Model No HPD-A20 HPD-A21 Wavelength range Single wavelength ( 190 to 800 nm ) Dual wavelength ( 190 to 800 nm ) Wavelength accuracy ± 0.5 nm ± 2 nm Wavelength repeatability 0.15 nm 0.3 nm Band width 8 nm 8 nm Optical path 10 nm 10 nm Deuterium lamp 190 to 400 nm 190 to 400 nm Tungsten lamp 400 to 800 nm 400 to 800 nm Baseline noise ( Static ) Au ≤ ± 0.35 × 10-5 ≤ ± 0.5 × 10-5 Baseline drift ( Static ) Au / h ≤ 1 × 10-4 ≤ 1.5 × 10-4 Minimum detection limit ≤ 4 × 10-8 ≤ 4 × 10-8 Power 100 W 100 W Power supply 90 V / 50 to 60 Hz 90 V / 50 to 60 Hz Dimension 460 × 315 × 180 mm 460 × 315 × 180 mm Weight 9.5 kg 9.5 kg |
Pump Specifications | Flow rate range: 0.0001 to 10 ml / min Maximum pressure: 50 MPa , 0 to 2 ml / min Column size: 4.6 × 250 mm Sample size: 5 µm Pressure pulsation: ≤ 1 % at 10 MPa Flow rate accuracy: ± 0.15 % at 1 ml / min Flow rate repeatability %: ≤ 0.1 Gradient accuracy %: ± 1.5 Gradient repeatability %: ≤ 0.2 Dimension: 460 × 310 × 180 mm Power supply: 90 V Weight: 11.6 kg |
Detector Specifications | Model No HPD-A20 HPD-A21 Wavelength range Single wavelength ( 190 to 800 nm ) Dual wavelength ( 190 to 800 nm ) Wavelength accuracy ± 0.5 nm ± 2 nm Wavelength repeatability 0.15 nm 0.3 nm Band width 8 nm 8 nm Optical path 10 nm 10 nm Deuterium lamp 190 to 400 nm 190 to 400 nm Tungsten lamp 400 to 800 nm 400 to 800 nm Baseline noise ( Static ) Au ≤ ± 0.35 × 10-5 ≤ ± 0.5 × 10-5 Baseline drift ( Static ) Au / h ≤ 1 × 10-4 ≤ 1.5 × 10-4 Minimum detection limit ≤ 4 × 10-8 ≤ 4 × 10-8 Power 100 W 100 W Power supply 90 V / 50 to 60 Hz 90 V / 50 to 60 Hz Dimension 460 × 315 × 180 mm 460 × 315 × 180 mm Weight 9.5 kg 9.5 kg |
Pump Specifications | Flow rate range: 0.0001 to 10 ml / min Maximum pressure: 50 MPa , 0 to 2 ml / min Column size: 4.6 × 250 mm Sample size: 5 µm Pressure pulsation: ≤ 1 % at 10 MPa Flow rate accuracy: ± 0.15 % at 1 ml / min Flow rate repeatability %: ≤ 0.1 Gradient accuracy %: ± 1.5 Gradient repeatability %: ≤ 0.2 Dimension: 460 × 310 × 180 mm Power supply: 90 V Weight: 11.6 kg |
Detector Specifications | Model No HPD-A20 HPD-A21 Wavelength range Single wavelength ( 190 to 800 nm ) Dual wavelength ( 190 to 800 nm ) Wavelength accuracy ± 0.5 nm ± 2 nm Wavelength repeatability 0.15 nm 0.3 nm Band width 8 nm 8 nm Optical path 10 nm 10 nm Deuterium lamp 190 to 400 nm 190 to 400 nm Tungsten lamp 400 to 800 nm 400 to 800 nm Baseline noise ( Static ) Au ≤ ± 0.35 × 10-5 ≤ ± 0.5 × 10-5 Baseline drift ( Static ) Au / h ≤ 1 × 10-4 ≤ 1.5 × 10-4 Minimum detection limit ≤ 4 × 10-8 ≤ 4 × 10-8 Power 100 W 100 W Power supply 90 V / 50 to 60 Hz 90 V / 50 to 60 Hz Dimension 460 × 315 × 180 mm 460 × 315 × 180 mm Weight 9.5 kg 9.5 kg |
Pump Specifications | Flow rate range: 0.0001 to 10 ml / min Maximum pressure: 50 MPa , 0 to 2 ml / min Column size: 4.6 × 250 mm Sample size: 5 µm Pressure pulsation: ≤ 1 % at 10 MPa Flow rate accuracy: ± 0.15 % at 1 ml / min Flow rate repeatability %: ≤ 0.1 Gradient accuracy %: ± 1.5 Gradient repeatability %: ≤ 0.2 Dimension: 460 × 310 × 180 mm Power supply: 90 V Weight: 11.6 kg |
Detector Specifications | Model No HPD-A20 HPD-A21 Wavelength range Single wavelength ( 190 to 800 nm ) Dual wavelength ( 190 to 800 nm ) Wavelength accuracy ± 0.5 nm ± 2 nm Wavelength repeatability 0.15 nm 0.3 nm Band width 8 nm 8 nm Optical path 10 nm 10 nm Deuterium lamp 190 to 400 nm 190 to 400 nm Tungsten lamp 400 to 800 nm 400 to 800 nm Baseline noise ( Static ) Au ≤ ± 0.35 × 10-5 ≤ ± 0.5 × 10-5 Baseline drift ( Static ) Au / h ≤ 1 × 10-4 ≤ 1.5 × 10-4 Minimum detection limit ≤ 4 × 10-8 ≤ 4 × 10-8 Power 100 W 100 W Power supply 90 V / 50 to 60 Hz 90 V / 50 to 60 Hz Dimension 460 × 315 × 180 mm 460 × 315 × 180 mm Weight 9.5 kg 9.5 kg |
Pump Specifications | Flow rate range: 0.0001 to 10 ml / min Maximum pressure: 50 MPa , 0 to 2 ml / min Column size: 4.6 × 250 mm Sample size: 5 µm Pressure pulsation: ≤ 1 % at 10 MPa Flow rate accuracy: ± 0.15 % at 1 ml / min Flow rate repeatability %: ≤ 0.1 Gradient accuracy %: ± 1.5 Gradient repeatability %: ≤ 0.2 Dimension: 460 × 310 × 180 mm Power supply: 90 V Weight: 11.6 kg |