Analytical HPLC System LAHS-B11 is a unit equipped with LCD display UV-Visible wavelength range of 190 nm to 700 nm and double plunger reciprocating pump system for constant flow pump. It has gradient kit system for test result accuracy. Consists of separate workstation, sample injector and front mounted self-locking switch. Deuterium lamp is used as light source with valve mounting frame. It is ideally used in the detection of compounds in pharmaceutical industries, research institute, food industries, and petroleum industries, etc.
Widely used in pharmaceutical, petroleum, research institute, food industrial applications in separation and detection of mixture of compounds in chemical analysis, poison analysis, environmental analysis, protein food detection, medicine testing, and polymer analysis.
High Pressure Constant Flow Pump |
Double plunger reciprocating pump: 2 sets Flow range: 0.001 to 10.00 mL/min Setting step size: 0.001 mL/min Flow accuracy: ≤ ± 0.3 % Flow stability: RSD ≤ 00.1 % Maximum pressure: 42 MPa Display pressure error: ≤ 1.0 % Pressure fluctuation: ≤ 1.0 % Power consumption: 80 W |
UV/Visible Detector |
Wavelength range: 190 to 700 nm Wavelength repeatability: ± 0.1 nm Band width: 8 nm Response time: 0 to 4.9 seconds Linearity range: 1.8 AU (5 %) Baseline Noise: ≤ ± 1 × 10-5AU Baseline drift: ≤ 2.0 × 10-4AU/h Minimum detectable concentration: 4.0 × 10-9 g/mL Detector cell: Optical path: 10 mm Volume: 10 µL Light source: Deuterium lamp Power consumption: 100 W Voltage: AC 110 V/ 220 V ± 10 %, 50/60 Hz Packing Dimensions (W × D × H): 600 × 440 × 380 mm 3 × (600 × 410 × 300 mm Gross weight: 75 kg |
High Pressure Constant Flow Pump |
Double plunger reciprocating pump: 2 sets Flow range: 0.001 to 10.00 mL/min Setting step size: 0.001 mL/min Flow accuracy: ≤ ± 0.3 % Flow stability: RSD ≤ 00.1 % Maximum pressure: 42 MPa Display pressure error: ≤ 1.0 % Pressure fluctuation: ≤ 1.0 % Power consumption: 80 W |
UV/Visible Detector |
Wavelength range: 190 to 700 nm Wavelength repeatability: ± 0.1 nm Band width: 8 nm Response time: 0 to 4.9 seconds Linearity range: 1.8 AU (5 %) Baseline Noise: ≤ ± 1 × 10-5AU Baseline drift: ≤ 2.0 × 10-4AU/h Minimum detectable concentration: 4.0 × 10-9 g/mL Detector cell: Optical path: 10 mm Volume: 10 µL Light source: Deuterium lamp Power consumption: 100 W Voltage: AC 110 V/ 220 V ± 10 %, 50/60 Hz Packing Dimensions (W × D × H): 600 × 440 × 380 mm 3 × (600 × 410 × 300 mm Gross weight: 75 kg |
High Pressure Constant Flow Pump |
Double plunger reciprocating pump: 2 sets Flow range: 0.001 to 10.00 mL/min Setting step size: 0.001 mL/min Flow accuracy: ≤ ± 0.3 % Flow stability: RSD ≤ 00.1 % Maximum pressure: 42 MPa Display pressure error: ≤ 1.0 % Pressure fluctuation: ≤ 1.0 % Power consumption: 80 W |
UV/Visible Detector |
Wavelength range: 190 to 700 nm Wavelength repeatability: ± 0.1 nm Band width: 8 nm Response time: 0 to 4.9 seconds Linearity range: 1.8 AU (5 %) Baseline Noise: ≤ ± 1 × 10-5AU Baseline drift: ≤ 2.0 × 10-4AU/h Minimum detectable concentration: 4.0 × 10-9 g/mL Detector cell: Optical path: 10 mm Volume: 10 µL Light source: Deuterium lamp Power consumption: 100 W Voltage: AC 110 V/ 220 V ± 10 %, 50/60 Hz Packing Dimensions (W × D × H): 600 × 440 × 380 mm 3 × (600 × 410 × 300 mm Gross weight: 75 kg |
High Pressure Constant Flow Pump |
Double plunger reciprocating pump: 2 sets Flow range: 0.001 to 10.00 mL/min Setting step size: 0.001 mL/min Flow accuracy: ≤ ± 0.3 % Flow stability: RSD ≤ 00.1 % Maximum pressure: 42 MPa Display pressure error: ≤ 1.0 % Pressure fluctuation: ≤ 1.0 % Power consumption: 80 W |
UV/Visible Detector |
Wavelength range: 190 to 700 nm Wavelength repeatability: ± 0.1 nm Band width: 8 nm Response time: 0 to 4.9 seconds Linearity range: 1.8 AU (5 %) Baseline Noise: ≤ ± 1 × 10-5AU Baseline drift: ≤ 2.0 × 10-4AU/h Minimum detectable concentration: 4.0 × 10-9 g/mL Detector cell: Optical path: 10 mm Volume: 10 µL Light source: Deuterium lamp Power consumption: 100 W Voltage: AC 110 V/ 220 V ± 10 %, 50/60 Hz Packing Dimensions (W × D × H): 600 × 440 × 380 mm 3 × (600 × 410 × 300 mm Gross weight: 75 kg |
High Pressure Constant Flow Pump |
Double plunger reciprocating pump: 2 sets Flow range: 0.001 to 10.00 mL/min Setting step size: 0.001 mL/min Flow accuracy: ≤ ± 0.3 % Flow stability: RSD ≤ 00.1 % Maximum pressure: 42 MPa Display pressure error: ≤ 1.0 % Pressure fluctuation: ≤ 1.0 % Power consumption: 80 W |
UV/Visible Detector |
Wavelength range: 190 to 700 nm Wavelength repeatability: ± 0.1 nm Band width: 8 nm Response time: 0 to 4.9 seconds Linearity range: 1.8 AU (5 %) Baseline Noise: ≤ ± 1 × 10-5AU Baseline drift: ≤ 2.0 × 10-4AU/h Minimum detectable concentration: 4.0 × 10-9 g/mL Detector cell: Optical path: 10 mm Volume: 10 µL Light source: Deuterium lamp Power consumption: 100 W Voltage: AC 110 V/ 220 V ± 10 %, 50/60 Hz Packing Dimensions (W × D × H): 600 × 440 × 380 mm 3 × (600 × 410 × 300 mm Gross weight: 75 kg |
High Pressure Constant Flow Pump | Double plunger reciprocating pump: 2 sets Flow range: 0.001 to 10.00 mL/min Setting step size: 0.001 mL/min Flow accuracy: ≤ ± 0.3 % Flow stability: RSD ≤ 00.1 % Maximum pressure: 42 MPa Display pressure error: ≤ 1.0 % Pressure fluctuation: ≤ 1.0 % Power consumption: 80 W |
UV/Visible Detector | Wavelength range: 190 to 700 nm Wavelength repeatability: ± 0.1 nm Band width: 8 nm Response time: 0 to 4.9 seconds Linearity range: 1.8 AU (5 %) Baseline Noise: ≤ ± 1 × 10-5AU Baseline drift: ≤ 2.0 × 10-4AU/h Minimum detectable concentration: 4.0 × 10-9 g/mL Detector cell: Optical path: 10 mm Volume: 10 µL Light source: Deuterium lamp Power consumption: 100 W Voltage: AC 110 V/ 220 V ± 10 %, 50/60 Hz Packing Dimensions (W × D × H): 600 × 440 × 380 mm 3 × (600 × 410 × 300 mm Gross weight: 75 kg |
High Pressure Constant Flow Pump | Double plunger reciprocating pump: 2 sets Flow range: 0.001 to 10.00 mL/min Setting step size: 0.001 mL/min Flow accuracy: ≤ ± 0.3 % Flow stability: RSD ≤ 00.1 % Maximum pressure: 42 MPa Display pressure error: ≤ 1.0 % Pressure fluctuation: ≤ 1.0 % Power consumption: 80 W |
UV/Visible Detector | Wavelength range: 190 to 700 nm Wavelength repeatability: ± 0.1 nm Band width: 8 nm Response time: 0 to 4.9 seconds Linearity range: 1.8 AU (5 %) Baseline Noise: ≤ ± 1 × 10-5AU Baseline drift: ≤ 2.0 × 10-4AU/h Minimum detectable concentration: 4.0 × 10-9 g/mL Detector cell: Optical path: 10 mm Volume: 10 µL Light source: Deuterium lamp Power consumption: 100 W Voltage: AC 110 V/ 220 V ± 10 %, 50/60 Hz Packing Dimensions (W × D × H): 600 × 440 × 380 mm 3 × (600 × 410 × 300 mm Gross weight: 75 kg |
High Pressure Constant Flow Pump | Double plunger reciprocating pump: 2 sets Flow range: 0.001 to 10.00 mL/min Setting step size: 0.001 mL/min Flow accuracy: ≤ ± 0.3 % Flow stability: RSD ≤ 00.1 % Maximum pressure: 42 MPa Display pressure error: ≤ 1.0 % Pressure fluctuation: ≤ 1.0 % Power consumption: 80 W |
UV/Visible Detector | Wavelength range: 190 to 700 nm Wavelength repeatability: ± 0.1 nm Band width: 8 nm Response time: 0 to 4.9 seconds Linearity range: 1.8 AU (5 %) Baseline Noise: ≤ ± 1 × 10-5AU Baseline drift: ≤ 2.0 × 10-4AU/h Minimum detectable concentration: 4.0 × 10-9 g/mL Detector cell: Optical path: 10 mm Volume: 10 µL Light source: Deuterium lamp Power consumption: 100 W Voltage: AC 110 V/ 220 V ± 10 %, 50/60 Hz Packing Dimensions (W × D × H): 600 × 440 × 380 mm 3 × (600 × 410 × 300 mm Gross weight: 75 kg |
High Pressure Constant Flow Pump | Double plunger reciprocating pump: 2 sets Flow range: 0.001 to 10.00 mL/min Setting step size: 0.001 mL/min Flow accuracy: ≤ ± 0.3 % Flow stability: RSD ≤ 00.1 % Maximum pressure: 42 MPa Display pressure error: ≤ 1.0 % Pressure fluctuation: ≤ 1.0 % Power consumption: 80 W |
UV/Visible Detector | Wavelength range: 190 to 700 nm Wavelength repeatability: ± 0.1 nm Band width: 8 nm Response time: 0 to 4.9 seconds Linearity range: 1.8 AU (5 %) Baseline Noise: ≤ ± 1 × 10-5AU Baseline drift: ≤ 2.0 × 10-4AU/h Minimum detectable concentration: 4.0 × 10-9 g/mL Detector cell: Optical path: 10 mm Volume: 10 µL Light source: Deuterium lamp Power consumption: 100 W Voltage: AC 110 V/ 220 V ± 10 %, 50/60 Hz Packing Dimensions (W × D × H): 600 × 440 × 380 mm 3 × (600 × 410 × 300 mm Gross weight: 75 kg |